Operating basics
PattXlm.SyncFrameEnd = 100;
PattYRe.Values = Seq3;
PattYRe.SyncFrameEnd = 100;
PattYlm.Values = Seq4;
PattYlm.SyncFrameEnd = 100;
The code assigns the user’s pattern variables Seq1, Seq2, Seq3, and Seq4 to the
four tributaries. These variables must be loaded into the separate MATLAB
Command Window as shown in the following
fi
gure.
In the case shown, a previously saved .mat
fi
le is loaded and the Seq variable is
created using the PattXRe.Values content. The
fi
gure also shows the resulting size
of the Seq variable as well as the
fi
rst 10 values. The pattern for each tributary
may have any length, but must be a row vector containing logical values.
Synchronizing a long pattern can take a long time. The easiest way to keep
calculations fast when using non-PRBS patterns longer than 2
15
, and if using
record lengths long enough to capture at least as many bits as in the pattern, is to
simply use the .SyncFrameEnd
fi
eld as shown above. Otherwise contact customer
support for help in optimizing the synchronization.
Example capturing unknown pattern.
Another way to load the pattern variable
when using a pattern that is not one of the PRBS selections is to use the OUI to
capture the pattern and store it in a variable. Do the following:
1.
Connect the optical signal with the desired modulation pattern to the OUI.
2.
Set up the Analysis Parameters properly with the exception of the data pattern
which is not yet known.
3.
Choose
Unknown
as the data pattern (do not choose “User Pattern” yet).
Optimize the signal to achieve open eye-diagrams and low EVM so that no
errors are expected.
40
OM4000D Series Coherent Lightwave Signal Analyzer
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...