Appendix E: Equivalent-Time (ET) oscilloscope operation
Prerequisites.
1.
System should be set up and de-skewed following the procedure above.
2.
Connect the Reference from Laser 2 to the Reference input with short
PM/APC jumper.
3.
Use a standard SM/APC (not PM)
fi
ber to connect the Laser 1 to the Signal
input.
4.
Use the LRCP to set the Signal 1 power level to about 10 dBm to start.
5.
Use the LRCP to turn on both lasers and tune them to the same channel where
you will be working. Set the Laser 1 power to get about 100mV pp.
6.
Use the LRCP to tune Laser 1 off grid by 100 to 500 MHz.
7.
Click
Run
on the oscilloscope; the data should look like randomly sampled
sine waves.
8.
Move the SM
fi
ber around until you get most signal on RX channels X-I and
X-Q (at least 3:1 ratio between X-I and Y-I. This is most easily done with a
polarization controller).
9.
Tape the
fi
ber down so that you continue to get most signal on X-I and X-Q.
10.
Click
Single
on the oscilloscope.
Procedure to inspect and correct the X-polarization calibration.
1.
Use the Optametra OUI to connect to the scope. Record length of ~8000
points recommended.
2.
Display the Matlab Engine Window, the X-Constellation Window and the
Y-Constellation Window. Close other windows.
3.
Be sure the Constellation Continuous traces is checked (selected) to show
the green curves.
4.
Put DispCalEllipsesET in the Matlab Engine Window of the OUI. Delete
everything else in the Matlab Engine Window.
5.
Click
Run
on the OUI.
6.
Observe that the ellipses are displayed on the Constellation plots. Right now
only the X-constellation has signal. The green trace should line up with the
blue circle in the X-constellation plot.
136
OM4000D Series Coherent Lightwave Signal Analyzer
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...