Appendix C: Calibration and adjustment (RT oscilloscope)
Click
Run
on the oscilloscope to get a rapidly updating display. You should
see 100 to 500 MHz sine waves.
Move the SM
fi
ber around until you get most signal on RX channels X-I and
X-Q (at least 3:1 ratio between X-I and Y-I. This is most easily done with a
polarization controller).
Tape the
fi
ber down so that you continue to get most signal on X-I and X-Q.
Click
Single
on the oscilloscope.
View and correct the
X-polarization calibration
(RT)
1.
Use the Optametra OUI to connect to the oscilloscope. Record length of
~20,000 points recommended in single block (BlockSize > 20,000).
2.
Display the MATLAB Engine Window, the X-Constellation Window and the
Y-Constellation Window. Close other windows.
3.
Put DispCalEllipses in the MATLAB Engine Window of the OUI. Delete
everything else in the MATLAB Engine Window.
4.
Click
Run
on the Optametra OUI.
5.
Observe that the ellipses are displayed on the Constellation plots. Right now
only the X-constellation has a signal. The green trace should line up with the
blue circle in the X-constellation plot.
6.
If the green trace in X-Constellation is elliptical:
Click
Stop
.
Enter CorrectX in the separate MATLAB Application Window.
Copy and paste the resulting pHyb statement before DispCalEllipses in the
MATLAB Engine window in the OUI as shown in the following image.
116
OM4000D Series Coherent Lightwave Signal Analyzer
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...