Appendix F: Con
fi
guring two Tektronix 70000 series oscilloscopes
The trigger holdoff is set to a
fi
xed time which is managed by the OUI. Longer
holdoffs are required for longer records.
In the Options tab, set the Master Scope Trigger Holdoff as required. Make
sure that the Slave Scope Trigger Holdoff is set to minimum time.
Slave oscilloscope trigger
settings
The slave uses the Reference signal from the master via the rear-panel BNC
The Ch 4 trigger settings should be identical to the master so that both take
data simultaneously. Residual skew will be handled by the OUI.
The slave scope has only one trigger. Select the default or minimum holdoff.
OM4000D Series Coherent Lightwave Signal Analyzer
149