Appendix E: Equivalent-Time (ET) oscilloscope operation
Oscilloscope sampling rate
Reference magnitude
response
reference phase
≤
2x the bandwidth of the
OM4000 (BW)
Flat
Linear
> 2x the bandwidth of the
OM4000 (BW)
4th order digital (bilinear)
Bessel
fi
lter with 3 dB cutoff
at BW + 2 GHz
Linear
Equalization is speci
fi
c to the sampling rate of the oscilloscope. As an example, if
the bandwidth of the OM4000 is 30 GHz and the sampling rate of the oscilloscope
is 80 Gs/s, the combined effect of the OM4000 receiver front end response and
the digital equalization
fi
lter will be that of a 4th order digital Bessel
fi
lter with
3 dB cutoff at 32 GHz, as shown in the following graph.
The digital
fi
lter is applied directly to the MATLAB workspace variables
Vblock(1).Values
through
Vblock(4).Values
using a 100 tap FIR
fi
lter.
Equalization provides a challenge when working in ET mode. Equivalent-Time
captured signals can be equalized to see the corrected waveshape when using
small enough time resolution or time step to emulate a real-time captured
waveform. A number of
fi
lters are provided with the most commonly used time
steps which are valid for correction.
The OUI searches through the Program Files\Optametra\OUI4006\ExecFiles
directory and applies the
fi
lter it
fi
nds with the proper time step. If the proper
fi
lter is not found, then none is applied and a warning is printed in the Engine
Command Response Window. When the SNR is < 20 dB in ET mode, it is best to
turn off the correction
fi
lter using EqFilt = []; in the Engine Command Window.
For more information, or for equalization support of a different scope sampling
rate, contact customer support.
OM4000D Series Coherent Lightwave Signal Analyzer
139
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...