Appendix E: Equivalent-Time (ET) oscilloscope operation
7.
If the green trace in X-Constellation is elliptical:
Click
Stop
.
Type
CorrectX
in the separate Matlab Application Window.
Copy and paste the resulting pHyb statement before DispCalEllipses in
the Matlab Engine Window in the OUI as shown below.
8.
Click
Run
. The green trace should now be circular. If it is not, repeat the
procedure one time.
Procedure to inspect and correct the Y-polarization calibration (ET).
1.
Move the input
fi
ber to get most of the signal on RX channels Y-I and Y-Q.
Tape it down.
2.
Click
Single
on the Oscilloscope.
3.
Click
Run
on the OUI.
4.
Observe that the ellipses are displayed on the Constellation plots. Right now
only the Y-constellation has signal. The green trace should line up with the
blue circle in the Y-constellation plot.
5.
If the green trace in Y-Constellation is elliptical:
Click
Stop
.
Type CorrectY in the separate Matlab Engine window.
Copy and paste the resulting pHyb statement before DispCalEllipses,
replacing any other pHyb statement.
6.
Click
Run
. The green trace should now be circular. If it is not, repeat this
procedure one time.
OM4000D Series Coherent Lightwave Signal Analyzer
137
Summary of Contents for OM4006D
Page 2: ......
Page 6: ......
Page 22: ...Compliance information xvi OM4000D Series Coherent Lightwave Signal Analyzer...
Page 24: ...Preface xviii OM4000D Series Coherent Lightwave Signal Analyzer...
Page 100: ...Taking measurements 76 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 146: ...Appendix D Automatic receiver deskew 122 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 202: ...Appendix H Cleaning and maintenance 178 OM4000D Series Coherent Lightwave Signal Analyzer...
Page 205: ...Index W Waveform averaging 48 OM4000D Series Coherent Lightwave Signal Analyzer 181...