Taking measurements
As each channel is analyzed, only that portion of the plot will be updated while
the most recent data displayed will continue to be shown in the other regions so
that an aggregate view of the multicarrier group can be displayed. Use the Clear
Data button to discard prior data.
Figure 10: Multicarrier constellation plots
Multicarrier Eye diagrams.
The Multicarrier Eye Diagrams are accessed by
clicking on the Eye icon button on the Home tab. These plots behave in a similar
fashion to the existing eye plots except that there are regions reserved for each
channel. The layout is controlled by the Multicarrier Setup window described
above.
As each channel is analyzed, only that portion of the plot will be updated while
the most recent data displayed will continue to be shown in the other regions so
that an aggregate view of the multicarrier group can be displayed. Use the Clear
Data button to discard prior data.
OM4000D Series Coherent Lightwave Signal Analyzer
73