Erasing oscilloscope memory
You can erase all setup and waveform information saved in the nonvolatile
oscilloscope memory with the TekSecure function. If you have acquired
confidential data on your oscilloscope, you may want to execute the TekSecure
function before you return the oscilloscope to general use. The TekSecure
function:
■
Replaces all waveforms in all reference memories with null values
■
Replaces the current front panel setup and all stored setups with the default
setup
■
Displays a confirmation or warning message, depending on whether the
verification is successful or unsuccessful
To use TekSecure:
1.
Push
Utility
.
2.
Push
Utility Page
.
Utility Page
2
3.
Turn multipurpose knob
a
and select
Security.
.
4.
Push
TekSecure Erase Memory
.
Utility Page
Security
TekSecure
Erase Memory
3
4
5.
Push
OK Erase Setup and Ref Memory
on the side menu.
OK
Erase Setup &
Ref Memory
5
To cancel the procedure, push
Menu Off
.
Save and recall information
116
MDO4000C Series Oscilloscope User Manual
Preliminary