CAN bus trigger
You can trigger on
Start of Frame
.
Type of Frame
,
Identifier
,
Data
,
Id &
Data
,
End of Frame
, and
Missing Ack
.
If you are setting up a CAN trigger and have made a
Trigger On
selection of
Type of Frame
, push
Frame Type
on the lower menu, and select
Data Frame
,
Remote Frame
,
Error Frame
, or
Overload Frame
.
If you have made a
Trigger On
selection of
Identifier
, push
Identifier
on the
lower menu and select a
Format
. Then push
Identifier
on the side menu, and
enter a binary or hex value with multipurpose knobs
a
and
b
.
Push
Direction
on the lower menu and select the direction of interest:
Read
,
Write
, or
Read or Write
.
If you have made a
Trigger On
selection of
Data
, push
Data
on the lower menu
and enter the parameters of interest.
LIN bus trigger
You can trigger on
Sync
,
Identifier
,
Data
,
Id & Data
,
Wakeup Frame
,
Sleep
Frame
, or
Error
.
If you are setting up an LIN trigger and have made a
Trigger On
selection of
Identifier
,
Data
, or
Identifier & Data
, push
Identifier
or
Data
on the lower
menu and enter the parameters of interest on the resulting side menu.
If you have made a
Trigger On
selection of
Error
, push
Error Type
on the
lower menu and enter the parameters of interest on the side menu.
Trigger setup
60
MDO4000C Series Oscilloscope User Manual
Preliminary
Summary of Contents for MDO4000C Series
Page 2: ...P r e l i m i n a r y...
Page 20: ...Compliance Information xvi MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...
Page 74: ...Acquire the signal 52 MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...
Page 140: ...Save and recall information 118 MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...