Measurement
Description
Mean
The arithmetic mean over the entire waveform or gated region.
Cycle Mean
The arithmetic mean over the first cycle in the waveform or the
first cycle in the gated region.
RMS
The true Root Mean Square voltage over the entire waveform or
gated region.
Cycle RMS
The true Root Mean Square voltage over the first cycle in the
waveform or the first cycle in the gated region.
Table 5: Miscellaneous measurements
Measurement
Description
Positive Pulse
Count
The number of positive pulses that rise above the mid reference
crossing in the waveform or gated region.
Negative Pulse
Count
The number of negative pulses that fall below the mid reference
crossing in the waveform or gated region.
Rising Edge
Count
The number of positive transitions from the low reference value
to the high reference value in the waveform or gated region.
Falling Edge
Count
The number of negative transitions from the high reference value
to the low reference value in the waveform or gated region.
Area
Area measurement is a voltage over time measurement. It
returns the area over the entire waveform or gated region in volt-
seconds. Area measured above ground is positive; area
measured below ground is negative.
Cycle Area
A voltage over time measurement. The measurement is the area
over the first cycle in the waveform or the first cycle in the gated
region expressed in volt-seconds. The area above the common
reference point is positive, and the area below the common
reference point is negative.
Analyze waveform or trace data
80
MDO4000C Series Oscilloscope User Manual
Preliminary
Summary of Contents for MDO4000C Series
Page 2: ...P r e l i m i n a r y...
Page 20: ...Compliance Information xvi MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...
Page 74: ...Acquire the signal 52 MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...
Page 140: ...Save and recall information 118 MDO4000C Series Oscilloscope User Manual P r e l i m i n a r y...