![Tektronix LTA7P Series Service Manual Download Page 77](http://html1.mh-extra.com/html/tektronix/lta7p-series/lta7p-series_service-manual_1078035077.webp)
Performance Verification: Ajustment/Verification Fixture
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
4- 29
Use the following procedure to verify the internal clock frequency.
Parameter tested
Internal clock frequency
Equipment
required
Frequency counter (item 8)
Prerequisites
Warm-up time: 30 minutes for adjustment/verification fixture and test
equipment
1.
Set the jumper positions as listed in the table.
Jumper
Jumper name
Jumper setting
J13
P6434 Deskew/probe setup
3-5 connected
J15
Clock selection
INT
J16
Deskew
2-3 and 4-5 connected
J17
Setup/hold select
HOLD
J19
Clock polarity select
Removed
2.
Using the frequency counter, measure the oscillator frequency at pin J1-2.
The frequency should match what is listed on the Calibration Data report.
3.
Record this measurement on the Calibration Data Report.
Verify Internal Clock
Frequency
Summary of Contents for LTA7P Series
Page 4: ......
Page 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 86: ......
Page 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 171: ......
Page 172: ......