Adjustment/Verification Fixture Adjustment
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
5- 21
9.
Connect the CH1 probe to J2 and center the rising edge on the center
horizontal graticule. Trigger on the CH2 probe connected to J1.
10.
Connect the CH1 probe to J4 (SU/HOLD CLK OUT) and verify for a 0
skew (
±
25 ps).
The following procedure is used to adjust the setup time.
Parameter tested
Setup Time Adjustment
Equipment
Required
Adjustment/verification fixture and power supply (item 2)
Oscilloscope (item 3)
Two 1 M
Ω
10X oscilloscope probe (item 4)
Shorting jumpers (item 14)
Prerequisites
Warm-up time: 30 minutes for adjustment/verification fixture and test
equipment
1.
Install the following jumpers:
Jumper
Jumper name
Jumper setting
J13
P6434 Deskew/probe setup
1-3 and 5-6 connected
J15
Clock selection
INT
J16
Deskew
1-2 and 4-5 connected
J17
Setup/hold select
SETUP
J19
Clock polarity select
Removed
1.
Set up the oscilloscope by pressing Setup, and then press Factory Setup, and
then press OK Confirm Factory Init to return the oscilloscope to default
conditions.
2.
Set up the oscilloscope as listed below.
a.
Set up the CH1 Vertical menu as follows:
H
Coupling
DC/50
Ω
H
Fine Scale
200 mV/div
H
Position
--3.32 div
b.
Set up the Horizontal menu as follows:
H
Time Base
Main
Setup Time Adjustment
Summary of Contents for LTA7P Series
Page 4: ......
Page 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 86: ......
Page 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 171: ......
Page 172: ......