Performance Verification: Ajustment/Verification Fixture
4- 24
TLA7Nx, TLA7Px, & TLA7Qx Logic Analyzer Module Service Manual
b.
Set up the Horizontal menu as follows:
H
Time Base
Main
H
Record Length
5000
H
Horizontal Scale
Main Scale @ 5 ns/div
c.
Set up the Trigger menu as follows:
H
Source
CH1
H
Coupling
DC
H
Slope
+
H
Level
700 mV
H
Mode
Normal
d.
Set up the Measure menu as follows:
H
Select Measurement for CH1
Frequency
H
Gating
Off
e.
Set up the Cursor menu as follows:
H
Function
Off
f.
Set up the Acquire menu as follows:
H
Acquisition Mode
Sample
H
Repetitive Signal
On
6.
Using the oscilloscope and the custom probe adapter shown in Figure 4--10
on page 4--26, verify that the output frequency at J1 pin-2 on the adjustment/
verification fixture is 105 MHz.
7.
Press the Run/Stop button to stop the acquisition.
8.
Disconnect the test equipment from the adjustment/verification fixture.
9.
This completes the functional verification procedures for the adjustment/ver-
ification fixture.
Summary of Contents for LTA7P Series
Page 4: ......
Page 14: ...Service Safety Summary x TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 22: ...Introduction xviii TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 86: ......
Page 148: ...Troubleshooting 6 40 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 154: ...Electrical Parts List 8 2 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 170: ...Mechanical Parts List 10 14 TLA7Nx TLA7Px TLA7Qx Logic Analyzer Module Service Manual...
Page 171: ......
Page 172: ......