Model 8010 High Power Device Test Fixture User's Manual
Section 2: Connecting instruments
8010-900-01 Rev. C / March 2017
2-11
Model 4200 and Model 4210-SMU connections
The Model 8010 includes internal protection circuitry for the Models 4200 and 4210-SMU. In the case
of a device under test (DUT) failure, this circuitry protects the SMUs from high voltage from the
Model 2657.
The LO terminal is connected to chassis ground once you connect the Model 4200 or 4210-SMU to
the Model 8010.
Do not convert triaxial cables to BNC cables. Using BNC cables will remove protection from
SourceMeter Instrument voltages and may result in instrument damage. You must use triaxial
cables.
Figure 7: Model 8010 to Model 4200 and Model 4210-SMU wiring diagram