Section 9: Maintenance
Model 8010 High Power Device Test Fixture User's Manual
9-2
8010-900-01 Rev. C / March 2017
Connector and socket cleaning
The performance of connectors and sockets can degrade because of dirt build-up or improper
handling. You can clean connectors and sockets with methanol-dipped cotton swabs.
After cleaning connectors and sockets, allow them to dry for at least one hour in a 50 °C low-humidity
environment.