Section 4: Off-state MOSFET characterization of a power MOSFET Model 8010 High Power Device Test Fixture User's Manual
4-2
8010-900-01 Rev. C / March 2017
Equipment needed for this example
•
One Model 8010 High Power Test Fixture with the Model 8010-DTB board installed on the high
voltage side of the test fixture (see
(on page 2-3) for more information)
•
One Model 2657 High Power System SourceMeter instrument
•
One Series 2600 System SourceMeter instrument
•
One power MOSFET in a TO-247 package
•
Cables to connect the Model 2657 to the Model 8010: Three Model CA-554 cable assemblies
•
Cables to connect the Series 2600 to the Model 8010: Three Model 7078-TRX cable assemblies
•
One CA-558-2 interlock cable
•
Model 2611 or 2612 only: Adapter to connect the SMU to the Model 8010: One Model 2600-
TRIAX triaxial adapter
•
Cables to connect instruments to the device in the Model 8010: Two Model CA-563 and two
Model CA-560-0 cables
•
One GPIB cable or Ethernet cable
•
One TSP-Link cable
Device connections
Connect instruments to the Model 8010
Refer to the appropriate figures below to connect your SourceMeter instrument to the Model 8010.
The ground wires must be attached to a known protective earth (safety ground) before
powering on instruments. Failure to attach the ground wires to a known protective earth
(safety ground) may result in electric shock.
The ground wires must be attached to a known protective earth (safety ground) before
powering on instruments. Failure to attach the ground wires to a known protective earth
(safety ground) may result in electric shock.