Section 2: Connecting instruments
Model 8010 High Power Device Test Fixture User's Manual
2-8
8010-900-01 Rev. C / March 2017
Model 2635 and Model 2636 connections
The Model 8010 includes internal protection circuitry for the Model 2635 and 2636 SMUs. In the case
of a device under test (DUT) failure, this circuitry protects the SMUs from high voltage from the Model
2657.
Do not convert triaxial cables to BNC cables. Using BNC cables will remove protection from
SourceMeter Instrument voltages and may result in instrument damage. You must use triaxial
cables.
Figure 4: Model 8010 to Model 2635 and Model 2636 rear panel connections