![Tektronix KEITHLEY 8010 User Manual Download Page 54](http://html1.mh-extra.com/html/tektronix/keithley-8010/keithley-8010_user-manual_1079031054.webp)
Model 8010 High Power Device Test Fixture User's Manual
Section 3: Working with devices and instruments
8010-900-01 Rev. C / March 2017
3-31
3. Connect the 8020-DP discharge probe ground cable to the protective earth (safety ground) screw
terminal on the bias tee.
Nothing else should be connected to the protective earth (safety ground) screw terminal. Do not
stack protective earth cables.
Figure 43: 8020-DP connected to bias tee bracket ground
Do not open and remove the Model 8010 cover until you have properly grounded the
discharge probe!