Section 7: High voltage capacitance measurements
Model 8010 High Power Device Test Fixture User's Manual
7-8
8010-900-01 Rev. C / March 2017
The following graphics show how to make open and short terminal connections. Make sure that you
do not have a device inserted in the test fixture device socket. For more information about open and
short compensation, refer to the
Optimizing cables and connections for high voltage CV
(on page 3-26) topic.
The next figure shows how to perform an open connection test without a device inserted:
Figure 70: Model 8010 and CVU-3K-KIT 3-terminal DUT open for Ciss and Coss connections
The next figure shows how to perform a short connection test without a device inserted and how to
install a short between the fixture terminals (that normally would be connected to the source and gate
of the FET):
Figure 71: Model 8010 and CVU-3K-KIT 3-terminal DUT short for Ciss connection