![Tektronix KEITHLEY 8010 User Manual Download Page 35](http://html1.mh-extra.com/html/tektronix/keithley-8010/keithley-8010_user-manual_1079031035.webp)
Section 3: Working with devices and instruments
Model 8010 High Power Device Test Fixture User's Manual
3-12
8010-900-01 Rev. C / March 2017
Optimizing connections for high current DC pulses
The Model 8010 is capable of carrying up to 100 A pulses from two Model 2651 SMU instruments.
You can connect two Model 2651 SMU instruments to the Model 8010, plus, the SMU instruments
can be connected in parallel. Note that the high current pulses require low resistance and low
inductance connections to minimize voltage drops and voltage errors in the measurements.
The series resistance of a cable is related to the thickness of the current-carrying conductor; To
minimize error due to excessive series resistance, use cables of sufficient thickness to carry the
current. To maximize accuracy of voltage measurements made in a high current test setup, use 4-
wire (remote) sense connections between the high current SMU instrument and the DUT.
Observe that inductance is related to the loop area between the HI and LO leads of the high current
conductor. The Keithley Model 2651-KIT cable assembles are designed to have low resistance
inductance and these cables are recommended whenever the Model 2651 High Power System
SourceMeter instrument is used.
Excessive inductance may slow the rise time of a pulse and result in inaccurate fast ADC readings on
the leading and falling edges of the pulse. For a series resistance error, take note of the maximum
voltage that the SMU can tolerate between the force and sense leads. Exceeding this voltage will
result in a voltage measurement error. Reduce the test current to stay within the force-sense voltage
limitation.
Using the guard terminal
Never connect the guard terminal from any instrument to the LO terminal of any instrument
in the Model 8010 or to the chassis. Connecting guard to LO can disable the high voltage
protection that is installed across the Model 4200 or Series 2600 SourceMeter
®
Instrument
connections. This may result in hazardous live voltages being present at the HI, SHI, SLO, or
LO terminals.