Specification— 2236 Service
Table 1-1
Electrical Characteristics
Characteristics
Performance Requirements
Supplemental Information
VERTICAL DEFLECTION SYSTEM
Deflection Factor
Range
2 mV per division to 5 V per division
in a 1-2-5 sequence
5 mV per division to 5 V per division
gain is adjusted with VOLTS/DIV switch
set to 10 mV per division.
2 mV per division gain is adjusted
with VOLTS/DIV switch set to 2 mV
per division
Accuracy
+ 15°C to +35°C
±2%.
0°C to +50°C
±3% .a
Range of VOLTS/DIV Variable Control
Continuously variable between
settings. Increases deflection factor
by at least 2.5 to 1.
Step Response
Rise Time
0°C to + 35°C
5 mV per Division to 5 V per
Division
3.5 ns or less.8
Rise time is calculated from the formula:
0.35
Bandwidth (—3 dB)
2 mV per Division
3.9 ns or less.8
+35°C to + 5 0 8C
5 mV per Division to 5 V per
Division
3.9 ns or less.8
2 mV per Division
4.4 ns or less.8
Aberrations
Positive-Going Step
2 mV per Division
+5% , -5 % , 5% p-p.
Measured with 5-division reference
signal, centered vertically, from a
50 fl source driving a 50 0 coaxial cable
terminated in 50 fi at the input connector
with the VOLTS/DIV Variable control in
the CAL detent.
5 mV per Division to 0.5 V
per Division
+4% , -4 % , 4% p-p.
1 V per Division to 5 V per
Division
+ 14%, -1 4 % , 14% p-p.8
Bandwidth ( - 3 dB)
0°C to +35°C
5 mV per Division to 5 V per
Division
Dc to at least 100 MHz.
Measured with a vertically centered
6-division reference signal from a 50 0
source driving a 50 Q coaxial cable
that is terminated in 50 0, both at
the input connector and at the probe
input, with the VOLTS/DIV Variable
control in the CAL detent.
2 mV per Division
Dc to at least 90 MHz.
+35°C to +50°C
5 mV per Division to 5 V per
Division.
Dc to at least 90 MHz.8
2 mV per Division
Dc to at least 80 MHz.8
•Performance Requirement not checked in Service Manual.
1-2
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......