Operating Instructions— 2236 Service
Fig. 2-5. Horizontal controls.
1-2-5 sequence. To obtain calibrated sweep speeds,
the A and B SEC/DIV Variable control must be in the
calibrated detent (fully clockwise).
A SEC/DIV
—The calibrated sweep speed is
shown between the two black lines on the clear
plastic skirt. This switch also selects the delay time
for delayed-sweep operation when used in con
junction with the B DELAY TIME POSITION
control.
B SEC/DIV
—The B Sweep speed is set by pulling
out the DLY'D SWEEP knob and rotating it clock
wise to a setting opposite the white line scribed on
the knob. The B Sweep circuit is used only for de
layed-sweep operation.
A and B SEC/DIV Variable Control
— Provides con
tinuously variable, uncalibrated A Sweep speeds to at
least 2.5 times the calibrated setting. It extends the
slowest sweep speed to at least 1.25 s per division.
(
20
) X10 Magnifier Switch
—To increase displayed sweep
speed by a factor of 10, pull out the A and B SEC/DIV
Variable knob. The fastest sweep speed can be ex
tended to 5 ns per division. Push in the A and B
SEC/DIV Variable knob to regain the XI sweep
speed.
( 2 n HORIZONTAL MODE Switch
— Three-position switch
determines the mode of operation for the horizontal
deflection system, and for frequency, period, width,
and totalize measurements.
A
— Horizontal deflection is provided by the A
Sweep generator at a sweep speed determined by
the A SEC/DIV switch setting. Nongated fre
quency, period, width, and totalize measurements
are made on the A Trigger signal.
ALT
— Alternates the horizontal displays between
the A Sweep (with an intensified zone) and the B
Delayed Sweep. The A Sweep speed is determined
by the setting of the A SEC/DIV switch. The B
Sweep speed and the length of the intensified zone
on the A Sweep are both determined by the B
SEC/DIV switch setting. Gated frequency, period,
width, and totalize measurements are made on the
B Trigger signal. The gate interval is defined by the
length of the intensified zone and is valid only when
B Sweep is triggered.
B
— Horizontal deflection is provided by the B
Sweep generator at a sweep speed determined by
2-5
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......