Performance Check Procedure— 2236 Service
e.
CHECK— Reading is between .99949 ns and
1.00051
ms
.
f. Disconnect the test equipment from the instrument.
6. Check Dc Volts Zero
a. Select the DCV function.
b.
Connect one DMM test lead to the — MULTIMETER
INPUTS connector and short the tip to the + input.
c. CHECK— Reading is between —.0001 V and .0001 V.
d. Disconnect the DMM test lead from the instrument.
7.
Check Dc Volts Accuracy
a.
Connect the dc volts calibrator via a female-to-dual
banana adapter, a 50-Q cable, and a female-to-dual banana
adapter to the MULTIMETER INPUTS connector.
b. CHECK— Reading is within the limits shown in Table
4-6 for each calibrator output voltage.
Table 4-6
Dc Voltage Readout Checks
Calibrator Dc
Voltage (V)
Display Readout
Limits (V)
0.4
.3995 to .4005
-0 .4
-.3 9 9 5 to -.4 0 0 5
4
3.995 to 4.005
- 4
-3 .9 9 5 to -4 .0 0 5
40
39.95 to 40.05
-4 0
-3 9 .9 5 to -4 0 .0 5
400
399.5 to 400.5
-4 0 0
-3 9 9 .5 to -4 0 0 .5
c. Disconnect the test equipment from the instrument.
8. Check Ac Volts Zero
a. Select the AC RMSV function.
b.
Connect one DMM test lead to the — MULTIMETER
INPUTS connector and short the tip to the + input.
c. CHECK— Reading is less than .0006 V.
d. Disconnect the test equipment from the instrument.
9. Check Ac Volts Accuracy
a.
Connect the ac volts calibrator via a female-to-dual
banana adapter, a 50-U cable, and a female-to-dual banana
adapter to the MULTIMETER INPUTS connector.
b.
CHECK— Reading is within the limits shown in Table
4-7 for each calibrator output voltage from 20 Hz to 20 kHz.
Table 4-7
Ac Voltage Readout Checks
Calibrator Ac
Display Readout
Voltage (V)
Limits (V)
0.4
.3954 to .4046
4
3.954 to 4.046
40
39.54 to 40.46
300
296.4 to 303.6
c. Disconnect the test equipment from the instrument.
10. Check Channel 1 Dc Volts
a. Select the CH 1 V function.
b. Set:
CH 1 VOLTS/DIV
50 mV
Channel 1 Input Coupling
GND
c. CHECK— Reading is between —.0012 V and .0012 V.
REV JAN 1985
4-17
d. Set the Channel 1 Input Coupling switch to DC.
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......