Performance Check Procedure—2236 Service
12. Check Ohms
a. Select the S) function.
b. Connect one DMM test lead to the instrument and
short the tip to the other DMM input.
c. CHECK— Reading is less than .20 i).
d. Disconnect the test lead from the instrument.
e. Connect the ohms calibrator via a female-to-dual
banana adapter, a 50-i) cable, and a female-to-dual banana
adapter to the MULTIMETER INPUTS connector.
NOTE
The connection resistance must be less than 200 mfl.
Compensate for any measurement error in the 10-
and 100-0 checks by noting the reading obtained
when the calibrator output is shorted and adjust the
readings accordingly. Alternately, use the 4-wire sens
ing capability of the calibrator to eliminate measure
ment error.
f. CHECK— Reading is within the limits shown in Table
4-12 for each calibrator output resistance.
Table 4-12
Resistance Readout Checks
Calibrator
Display Readout
Resistance (ohm)
Limits (ohm)
10
9.77 to 10.23
100
99.6 to 100.4
1 k
.996 k to 1.004 k
10 k
9.96 k to 10.04 k
100 k
99.6 k to 100.4 k
1 M
.996 M to 1.004 M
10 M
9.96 M to 10.04 M
g. Disconnect the test equipment from the instrument.
NOTE
The next two checks use two high-resistance preci
sion resistors. To preserve the accuracy of these re
sistors, store them in a dean, dry environment and do
not allow finger contact or other forms of contamina
tion on the resistor body, if contamination is sus
pected, wash the resistor body with isopropyl alcohol
and air dry. if exposed to excessive humidity, bake
the resistors at 200° F for 4 hours.
h. Connect the DMM test leads to the instrument; then
connect the 150 Mi) precision resistor to the test leads.
i. CHECK— Reading is between
148.4 Mi) and
151.6 MS).
j. Disconnect the 150 MS) resistor and connect the
1.5 GS) precision resistor to the test leads.
k. CHECK— Reading is between 1.34 GS) and 1.66 GS).
I. Disconnect the test equipment from the instrument.
13. Check Diode Forward Drop
a. Select UPPER FUNCTIONS and push in the FREQ,
PER, and WIDTH buttons simultaneously to enter the self
test mode.
b. Push and release the SGL SWP button until S6 is
displayed.
c. Select the DCV function.
d. Set the Channel 1 Input Coupling switch to GND.
e. Connect the dc volts calibrator via a female-to-dual
banana adapter, a 50-S) cable, and a female-to-dual banana
adapter to the MULTIMETER INPUTS connector.
f. Set the calibrator to produce a 2.5-V output.
g. CHECK— Reading is Fd 2.474 V to 2.526 V.
h. Disconnect the test equipment from the instrument.
4-19
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......