Specification— 2236 Service
Table 1-2
Environmental Characteristics
Characteristics
Description
Temperature
NOTE
The instrument meets the requirements of MIL-T-28800C, paragraphs 4.5.5.1.3,
4.5.5.1.4, and 4.5.5.1.2.2 for Type III, Class 5 equipment, except for maximum
operating temperature, and minimum storage temperature as indicated.
Operating
AC RMSV, DCV, and Q Modes
0° to 40°C (+ 3 2 °F to +104°F).
All Other Modes
0° to +50°C (+ 3 2 ° to +122°F).
Nonoperating
—50°C to +75°C (—58°F to +167°F). Tested to MIL-T-28800C paragraphs
4.5.5.1.3 and 4.5.5.1.4, except in 4.5.5.1.3 steps 4 and 5 (0°C operating test) are
performed ahead of step 2 ( —50 °C nonoperating test). Equipment shall remain
off upon return to room ambient during step 6. Excessive condensation shall be
removed before operating during step 7.
Altitude
Operating
To 4,500 m (15,000 ft). Maximum operating temperature decreased 1 °C per
1,000 ft above 5,000 ft.
Nonoperating
To 15,000 m (50,000 ft).
Humidity (Operating and Nonoperating)
5 cycles (120 hours) referenced to MIL-T-28800C paragraph 4.5.5.1.2.2, for Type
III, Class 5 instruments. Nonoperating and operating of 95% - 5 % to +0%
relative humidity. Operating at +30°C and +40°C for AC RMSV, DCV, and fi
Modes only and operating at +30°C and +50°C for all other modes.
Nonoperating at +30°C to +60°C.
Vibration (Operating)
15 minutes along each of 3 major axes at a total displacementof 0.015 inch p-p
(2.4 g’s at 55 Hz) with frequency varied from 10 Hz to 55 Hz to 10 Hz in 1-minute
sweeps. Hold for 10 minutes at 55 Hz in each of the 3 major axes. All major
resonances must be above 55 Hz.
Shock (Operating and Nonoperating)
30 g’s, half-sine, 11-ms duration, 3 shocks per axis each direction, for a total of
18 shocks.
EMI
Meets radiated and conducted emission requirements per VDE 0871 Class B.
1-16
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......