VERTICAL
Performance Check Procedure— 2236 Service
Equipment Required (see Table 4-1):
Calibration Generator (Item 1)
Dual-Input Coupler (Item 10)
Leveled Sine-Wave Generator (Item 2)
10X Attenuator (Item 11)
Two 50-fi BNC Cables (Item 8)
BNC T-Connector (Item 12)
50-n BNC Termination (Item 9)
10X Probe (provided with instrument)
INITIAL CONTROL SETTINGS
Vertical (Both Channels)
POSITION
Midrange
VERTICAL MODE
CH 1
BW LIMIT
On (button in)
VOLTS/DIV
2 mV
VOLTS/DIV Variable
CAL detent
INVERT
Off (button out)
Input Coupling
DC
Horizontal
POSITION
Midrange
HORIZONTAL MODE
A
A SEC/DIV
0.2 ms
SEC/DIV Variable
CAL detent
X I0 Magnifier
Off (knob in)
A TRIGGER
VAR HOLDOFF
NORM
Mode
P-P AUTO
SLOPE
OUT
LEVEL
Midrange
A&B INT
VERT MODE
A SOURCE
INT
A EXT COUPLING
AC
PROCEDURE STEPS
1. Check Deflection Accuracy and Variable Range
a. Connect the standard-amplitude generator output via
a 50-fl cable to the CH 1 OR X input connector.
b. CHECK— Deflection accuracy is within the limits
given in Table 4-2 for each CH 1 VOLTS/DIV switch setting
and corresponding standard-amplitude signal. When at the
20-mV VOLTS/DIV switch setting, rotate the CH 1 VOLTS/
DIV Variable control fully counterclockwise and CHECK that
the display decreases to 2 divisions or less. Then return the
CH 1 VOLTS/DIV Variable control to the CAL detent and
continue with the 50-mV check.
c.
Move the cable from the CH 1 OR X input connector
to the CH 2 OR Y input connector. Set the VERTICAL
MODE switch to CH 2.
Table 4-2
Deflection Accuracy Limits
VOLTS/DIV
Standard
Vertical
Accuracy
Switch
Amplitude
Deflection
Limits
Setting
Signal
(Divisions)
(Divisions)
2 mV
10 mV
5
4.90 to 5.10
5 mV
20 mV
4
3.92 to 4.08
10 mV
50 mV
5
4.90 to 5.10
20 mV
0.1 V
5
4.90 to 5.10
50 mV
0.2 V
4
3.92 to 4.08
0.1 V
0.5 V
5
4.90 to 5.10
0.2 V
1 V
5
4.90 to 5.10
0.5 V
2 V
4
3.92 to 4.08
1 V
5 V
5
4.90 to 5.10
2 V
10 V
5
4.90 to 5.10
5 V
20 V
4
3.92 to 4.08
d. Repeat part b using the Channel 2 controls.
2.
Check Position Range
a. Set:
VOLTS/DIV (both)
50 mV
Input Coupling (both)
AC
b. Set the generator to produce a 0.5-V standard-
amplitude signal.
4-5
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......