Operating Instructions— 2236 Service
CH 1 V— Measures dc voltage or true ac rms volt
age signals applied to the CH 1 OR X & DMM verti
cal input connector. For dc voltage measurements
set the Channel 1 Input Coupling switch to DCV
position, and for true ac rms voltage measure
ments set the Channel 1 Input Coupling switch to
the AC RMSV position. There is no autoranging;
CH 1 VOLTS/DIV switch selects the range for the
CTM voltage measurements. The readout will indi
cate the measuring mode the CTM is in by display
ing an “Ac” or “dc” on the left side of the voltage
reading. In overrange condition the readout will dis
play the word “OUCH”.
In CH 1 V mode, the CTM has the capability of
storing a relative reference value, such that the dis
played value is the input voltage minus the stored
reference value. When either Channel 1 AC RMSV
or DCV volts are first entered, the stored reference
value is zero, and the Channel 1 input voltage is
displayed directly. To set the reference value,
press in the SGL SWP RESET button momen
tarily, and the value on the readout becomes the
stored reference value. To regain normal voltage
measurement press again the SGL SWP RESET
button momentarily or change the CH
1
VOLTS/DIV switch position.
AC RMSV— Measures the true ac rms value of the
input signal applied to the Multimeter inputs, from
0 V to 349.9 V rms. The word "OUCH” is displayed
if 350 V or more is applied to the Multimeter input
connector.
TEMP— Measures temperature in degrees Celsius
°C when the TEMP button is pressed in, or de
grees Fahrenheit °F when both TEMP and A TIME
buttons are pressed in. Temperature readings fol
lowing by °C or °F are displayed on the CTM
readout.
<=3)))<50 (Continuity)— Press in both AC RMSV
and TEMP buttons for continuity measurements.
For resistance readings below 5.012 the word
“Short” is displayed and an audible tone will be
generated. For resistance readings greater than or
equal to 5.012, the word “OPEN” is displayed.
Self Test Routine— The CTM can be entered into
a Self Test routine by setting the UPPER FUNC-
TIONS-LOWER FUNCTIONS switch to IN and
pressing in the FREQ, PER, and WIDTH buttons at
the same time. The message “SELF tESt” will be
displayed on the readout to indicate that the rou
tine is in self test mode. To exit from the routine
press in any of the CTM front panel buttons to re
gain normal measurement mode. Repeated press
ing of the SGL SWP RESET button will cause the
CTM to advance through the test menu. To exit
from the test menu, keep pressing the SGL SWP
RESET button until “End S-t” message is dis
played on the readout. Then exit from the Self Test
routine by pressing in any one of the CTM front
panel buttons. Self Test routine should normally be
performed by a qualified service personnal.
(
37
)
A TIME POSITION Control— Used in conjunction with
the B DELAY TIME POSITION control in A TIME
mode. The control determines the time difference be
tween the starts of the two intensified zones.
(3 8 ) Readout— Consists of a nine-digit vacuum fluores
cent unit which is used to display measurements se
lected by the CTM. No polarity indication is displayed
for positive values. Negative polarity indication is
automatic for negative values. In Counter and Timer
modes the decimal point is always placed in one of the
three most-significant digits, and the exponent is al
ways an integer multiple of 3. Resistance measure
ments are also displayed with an exponent. All
frequency measurements are in Hertz, and time mea
surements are in seconds.
RIGHT-SIDE PANEL
Refer to Figure 2-8 for location of item 39.
39)
Multimeter Connectors— Two banana like jacks pro
vide positive (red) and negative (black) inputs for volt
ages, resistance, and temperature measurements.
REAR PANEL
Refer to Figure 2-9 for location of item 40.
(4 0 ) EXT Z-AXIS Connector— Provides a means of con
necting external signals to the Z-axis amplifier to in
tensity modulate the crt. Applied signals do not affect
display waveshape. Signals with fast rise times and
fall times provide the most abrupt intensity change,
and a 5-V p-p signal will produce noticeable modula
tion. The Z-axis signals must be time-related to the
display to obtain a stable presentation on the crt.
2-9
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......