Operating Instructions— 2236 Service
no b trig— When either the A TIME or DLY TIME
mode and either ALT or B HORIZONTAL MODE is
selected with the A Sweep triggered and the B Trigger
circuit not receiving a trigger signal in delay mode, the
readout will display “no b trig”. This message is also
displayed with the instrument in either ALT or B HOR
IZONTAL MODE and either the FREQ or PER mode
selected, and the CTM is not receiving a trigger signal.
OPEN—The readout will display “OPEN” with the
CTM in either Ohms or Continuity mode (AC RMSV
and TEMP buttons pressed in) under the following
conditions; in Si mode one or both test leads discon
nected or when resistance exceeds 1.99 GSl, in Con
tinuity mode when the resistance equals or exceeds
5.0 SI.
diodE— In Resistance/Semiconductor mode the read
out will momentarily display “diodE”, and an audible
tone will be generated if the device being measured is
a forward-biased semiconductor junction. After ap
proximately one second, it will display the forward
voltage drop of the device.
Fd— In Resistance/Semiconductor mode “Fd” will be
displayed on the left side of the readout to indicate
that a forward voltage drop is being displayed.
Probe-?— When the temperature probe exceeds its
internal resistance limits the readout will display
“ProbE-?” to indicate that the temperature probe is
either faulty or disconnected from the Multimeter input
connectors.
Short— In Continuity mode the readout will display
“Short” when the measured resistance is less than
5.0 Si.
OUCH—The word “OUCH” is displayed when the in
put voltage to the multimeter inputs is 500 volts or
more in DCV mode, and 350 volts or more in AC
RMSV mode. In CH 1 V mode, the word ‘ OUCH’ is
displayed if overranged.
dc— In CH 1 V mode with Channel 1 Input Coupling
switch in GND or DCV position, “dc” will be displayed
on the left side of the readout.
Ac— In CH 1 V mode with Channel 1 Input Coupling
switch in AC RMSV position, “Ac” will be displayed on
the left side of the readout.
r— In CH 1 V Relative Reference mode, “r” will be
displayed on the right side of the readout.
no dELTA— When in A TIME mode with the measure
ment intensified zone (moved by rotating the A TIME
position control) dialed off the right end of the trace,
the readout will display “no dELTA”.
2-13
Summary of Contents for 2236
Page 10: ...2236 Service viii The 2236 Oscilloscope ...
Page 74: ...Theory of Operation 2236 Service 3 30 ...
Page 102: ...Performance Check Procedure 2236 Service 4 21 Figure 4 1 Test setup for DMM common mode check ...
Page 242: ...2236 Service 4204 34 Figure 9 6 CTM basic block diagram ...
Page 243: ...DMM POWER S U P P LIE S 2236 Service F ig u re 9 7 C T M d e ta ile d b lo c k diag ram ...
Page 254: ...S I 0 O I 5 1 S O i 4204 21 C IR C U IT BOARD INTERCONNECTIONS ...
Page 267: ...1 2 3 4 5 6 7 8 9 J T K L M_ _ ___ N___l _ P____l S ...
Page 271: ...1 2 3 4 5 6 7 8 9 lO P2SO O U1 A TRIGGER V T O W2500 5 A 223 R E V A P R W G ...
Page 275: ......
Page 280: ......
Page 285: ...A 6 F IL T E R B O A R D lo c a te d u n d e r H V S h ie ld ...
Page 313: ...2236 Service T O U I 3 0 0 4 8 K T O UI300 II S K 0 2 2 3 6 Z Z O A Z 2 OPTION IA ...
Page 317: ...A 10 C O U N T E R T IM E R M U L T IM E T E R B O A R D A D J U S T M E N T LO C A T IO N S ...
Page 319: ...2236 Service TROUBLESHOOTING GUIDE 4206 99 4204 69A ...
Page 323: ...9 ...
Page 324: ...2 REV APR 1986 ...
Page 325: ...2236 SERVICE ...
Page 326: ......
Page 332: ......
Page 333: ...2236 SERVICE ...
Page 334: ...2 4 2 1 ...
Page 338: ......