1. Performance Tests – Verifying Specifications
Equipment Required
3
Table 1-1. Recommended Test Equipment (Continued)
Digital Multimeter
Input Resistance: 10 M
Ω
Accuracy: 0.04% at 1 Vdc
T HP/Agilent
34401A
40kV ac/dc High
Voltage Probe
Range: >= 5KV dc, input Z= 1 G
Ω
,
Accuracy: 2%, 10 M
Ω
load
T HP/Agilent
34300A
Clip-on Current Probe
Range: >= 1 Amp dc
Output: 1 Vdc at 10 Amp dc
T HP/Agilent
34302A
Universal Counter
Range: >= 500 MHz
Accuracy: 0.1 PPM
T HP/Agilent
53131A,
Opt 010, 030
CW Microwave
Counter
Range: >= 10 GHz
Accuracy: >= 0.1 ppm
T HP/Agilent
5350B,
Option 010
Spectrum Analyzer
Frequency Range: >= 9.5 GHz
Resolution: 100 kHz / Division
Center Frequency Accuracy: 1 PPM
Input: Up to 1 watt input capability
Dynamic Range: 100 dB
T HP/Agilent
8566A/B
Input Probe
10:1 Divider (included with
54600A oscilloscope)
T HP/Agilent
10071A
Input Probe
1:1 1 M
Ω
(for 50
Ω
measurements)
T
HP/Agilent 10070A
RF Signal Generator
Frequency Range: > 80 MHz
Power: >= +7 dBm
T HP/Agilent
8656B
Pulse Generator
Repetition Rate: 10 Hz
Pulse Width: <= 10 usec
Output Voltage: >= +3V
T HP/Agilent
8111A
General-purpose DC
Power Supply,
Adjustable
0 - 10V,
>= 25 ma
T HP/Agilent
E3610A
IBM-Compatible PC
MS-DOS
®
3.3 or later, 640 kB RAM,
1 serial (COM: ) Port
T Any
Vendor
Service Accessories
Kit
Special Test Tools, Devices, and
Fixtures
T 05071-67003
Profiling Software
For Online CBT Performance Testing
(See Title Page for download
instructions)
T 05071-13501
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...