2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
114
3.7.2.1 Electron Multiplier High-Voltage Supply Test
1
Remove instrument power, and wait one minute for the A16 module voltages
to bleed down.
2
Re-connect the Ion Pump high-voltage supply cable to A17.
3
Disconnect the left-hand high-voltage lead and reconnect it to the tip of the
high-voltage probe using a short piece of insulated wire between the inside of
the high-voltage lead connector and the
high-voltage probe.
4
Ground the high voltage probe return to the 5071A chassis, and plug the
probe output into the DMM. Set voltage mode and at least 10.0 Volts range.
The input impedance of the DMM must equal that specified for the high-
voltage probe.
5
Apply instrument power, wait for the instrument to begin warming up, and
observe the voltage displayed by the DMM. Multiply this voltage by the
voltage-division ratio of the probe and record results.
6
Observe the voltage displayed on the LCD using the
INFO/EMULT
menu
selection. Note that the LCD does not display the sign of the voltage.
QUESTION:
Do the voltage readings agree to within 5%?
If Yes:
The A16 High-Voltage Supply Module is functional. Remove
the extender, replace A2. Re-connect A19 to A1 at A1J11, and
reconnect the semi-rigid cable from the SMA connector on A19
to A9J5. Continue testing.
If No:
Go to, 3.7.2.2.
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...
Page 197: ......