2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
88
A2-Mass-Spectrometer Diagnostic Tree
This functional subsection describes the A2 mass-spectrometer diagnostic
tree for the 5071A Primary Frequency Standard. Performance of these tests
is indicated by the presence of either a
self-test or fatal error that reports a mass-spectrometer fault.
3.1.3.1.0 Mass-Spectrometer Resistance Test
1
Remove instrument power.
2
Unplug CBT cables from A2J1 and A2J2.
3
Use a DMM to measure the resistance across A2P4 CBT cable
pins 1 and 4.
QUESTION:
Is the resistance greater than 200 k
Ω
?
If Yes:
Go to, 3.1.3.1.1.
If No:
Recheck results and DMM, then replace CBT.
3.1.3.1.1 Mass-Spectrometer Test (Current Off)
1
Remove instrument power.
2
Replace A2 CBT Controller with long extender board (05071-60052).
(Do not insert A2 at this time.)
3
Reapply instrument power: instrument will go to
Fatal Error
.
4
Use SCPI command
syst:print?
.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION: Is
Mass spec:
between -0.5 and +0.5V?
If Yes:
Go to, 3.1.3.1.2.
If No:
Remove power, replace A7, and retest.
3.1.3.1.2 Mass-Spectrometer Monitor Test (+ Current On)
1
Place a 10 k
Ω
resistor between pins 1 and 37 on connector XA2P2
(48-pin connector on extender card).
2
Use SCPI command
syst:print?
.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION: Is
Mass spec:
between 9.5 and 11V?
If Yes:
Go to, 3.1.3.1.3.
If No:
Remove power, replace A7, and retest.
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...
Page 197: ......