1
1. Performance Tests – Verifying
Specifications
Introduction
NOTE
In the interest of PRESERVING its usability the information in
this chapter has been retained in its original form without
changes. No attempt has been made to update the equipment,
accessories, or parts to current availability. It is therefore
necessary for the user of this manual to consider the
recommendations of test equipment and accessories as
suggestions. Some or all of these items may no longer be
available from Symmetricom or any other vendors. An updated
list of available replacement parts from Symmetricom is provided
in Chapter 5.
This chapter provides procedures to test the electrical performance of the
5071A Primary Frequency Standard, using the specifications listed in
Chapter 6, “Specifications.” Three types of tests are provided:
•
Operational
Verification
•
Complete Performance Tests
•
RS-232
Verification
Operational Verification
The Operational Verification test is an abbreviated series of checks that may
be performed to give a high degree of confidence that the instrument is
operating properly without performing the complete Performance Tests. An
operational verification is useful for incoming inspection, routine
maintenance, and after instrument repair.
Complete Performance Tests
The complete Performance Tests verify the specifications listed in Chapter 6
of this manual. All tests can be performed without access to the inside of the
instrument.
RS-232 Verification
The RS-232 Serial Port verification checks the serial port for successful data
communication with a terminal or PC.
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...
Page 197: ......