2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
50
FIRST WORD (Rightmost word):
-
0 0 0 0 0 0 0 0
⇐
Hex digit
/ \ / \ / \ / \ / \ / \ / \ / \
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
⇐
Binary digit
| | | | | | | | | | | | | | | \_1. Servo offset - No Message
| | | | | | | | | | | | | | \_2. CBT shutdown: <#> - Fatal Error
| | | | | | | | | | | | | \_3. VCXO near tuning limit - Warning
| | | | | | | | | | | | \_4. VCXO at tuning limit - Fatal Error
| | | | | | | | | | | \_5. C-field current at range limit - Fatal Error
| | | | | | | | | | \_6. E-multiplier voltage at range limit - Message
| | | | | | | | | \_7. Signal gain at range limit - Warning
| | | | | | | | \_8. Internal DC supply out of range - Warning
| | | | | | | \_9. Mass spec supply regulation failure - Fatal Error
| | | | | | \_10. Hotwire supply regulation failure - Fatal Error
| | | | | \_11. CBT oven supply regulation failure - Fatal Error
| | | | \_12. CBT oven voltage at range limit - Fatal Error
| | | \_13. RF amplitude at range limit - Warning
| | \_14. Microprocessor PLL unlocked - Warning
| \_15. 87 MHz PLL unlocked - Message
\_16. Unused bit - always zero.
SECOND WORD:
0 0 0 0 0 0 0 0
⇐
Hex digit
/ \ / \ / \ / \ / \ / \ / \ / \
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00
⇐
Binary digit
| | | | | | | | \
_17. DRO unlocked - Message
| | | | | | | \
_18. VCXO oven error - Message
| | | | | | \
_19. CBT signal loss - Fatal Error
| | | | | \
_20. CBT signal low - Warning
| | | | \
_21. Servo system is in manual mode - Warning
| | | \
_22. Attempted lock on wrong peak - Fatal Error
| | \
_23. Center peak saturated - Message
\-----
reserved
--------/
Example Error-Code Interpretation
-
You can create a fatal error by disconnecting A5P2 from A5 after the instrument has
powered-up and is running (wait up to 35 seconds for the software to confirm this
failure). When the attention LED begins flashing, you're ready to read the error code
from the log.
-
Typical log display:
Error code: 000000a0 20000000
-
By using the above error-message map, the following information may be decoded:
-
15. 87 MHz PLL unlocked - Message
-
19. CBT signal loss - Fatal Error (This is the event that initiated the shut-down.)
-
20. CBT signal low – Warning
NOTE
You must power-off, reconnect A5P2, and then power-on again
to clear this error
.
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...
Page 197: ......