2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
56
1.3.8 RF-Level Control-Current Test
The amplitude of the microwave probe signal is controlled by an amplitude
modulator in A15 driven with current generated from a DAC on A7. If control
current is absent, amplitude modulation is lost.
1
Remove power and disconnect the cable from A1J19 to A15J1.
2
Using a suitable SMA adapter, connect the DMM to the cable from A1J19
and set the DMM to measure at least 16 mA dc.
3
Disconnect the CBT from A2 at A2J2. Cycle instrument power and wait for it
to enter the
Fatal Error
mode.
4
Observe and record the current measured at A1J19.
5
Remove power and reconnect the CBT cable to A2J2.
6
Reapply power to the instrument and verify that LCD displays
Warming up
.
7
Quickly observe and record the average current at A1J19.
8
Remove power and reconnect the cable to A1J19.
QUESTION: Was
the
Fatal Error
current 15.8
±
0.3 mA, and the
Warming up
current 1.58
±
0.03 mA?
If Yes:
The control current circuit is operational.
If No:
Remove power, replace A7, and retest.