2. Service
Assembly/Module Diagnostic Trees (Diagnostic Section 3)
87
3.1.2.1.3 Thermistor Monitor Test 1 (Current On)
1
Place a 10 k
Ω
resistor between pins 1 and 7 on connector XA2P2
(48-pin connector on extender card).
2
Use SCPI command
syst:print?
.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION:
Is thermistor between 9.35 and 9.75 C
°
?
If Yes:
Go to, 3.1.2.1.4.
If No:
Remove power, replace A7, and retest.
3.1.2.1.4 Thermistor Monitor Test 2 (Current On)
1
Place a 10 k
Ω
resistor between pins 3 and 7 on connector XA2P2.
2
Use SCPI command
syst:print?
.
(Terminal-to-instrument via RS-232C data communication required.)
QUESTION:
Is thermistor between -9.35 and -9.75 C
°
?
If Yes:
Remove power, replace A2, and retest.
If No:
Remove power, replace A7, and retest.