2. Service
Top-Level Diagnostic Tree (Diagnostic Section 1)
53
1.3.3 CBT-Signal Fatal Errors
These errors occur when the instrument completes its lock-up procedure, and
the frequency servo loop cannot operate correctly. They usually indicate the
CBT output signal is unstable. This can be due to a malfunction in the CBT
itself, or more frequently, in the circuits associated with it. The following
procedure tests these alternatives.
1
Perform the Analog-Signal Chain tests (2.2.1.0) and repair as required.
2
Perform the RF-Chain tests (2.3.1.0) and repair as required.
3
Perform the A16 Electron-Multiplier Supply tests (3.7.2.1) and repair as
required.
4
Perform the RF-Level Control-Current Check, (1.3.8), and repair as required.
5
Apply power to the instrument: observe the warm up and subsequent
operation.
QUESTION:
Did the instrument fail in the same manner?
If Yes:
The CBT is possibly defective, contact Symmetricom for further
assistance.
If No:
Problem may be resolved - retest.
1.3.4 CBT Shutdown Message
1
Read the CBT shutdown message preceding the error code.
QUESTION:
What is the message?
If
CBT shutdown:
thermistor open
, Go to, A2 thermistor
(3.1.2.1.0) diagnostic tree.
If
CBT shutdown:
thermistor shorted
, Go to, A2 thermistor
(3.1.2.1.0) diagnostic tree.
If
CBT shutdown:
ion pump overcurrent
, Go to, Excess Ion-
pump current (1.3.2.1) test.
If CBT shutdown:
E-mult adjust failed – This indicates that although
larger e-mult movements were attempted the internal CBT parameters
could not be compensated sufficiently to allow continued operation.
Repair is likely necessary, however a re-start of the 5071A should first
be attempted as a possible remedy. If the failure re-occurs with the
same message, contact Symmetricom to arrange repair/replacement of
the CBT. If the error is not repeated but another error occurs, follow the
procedures for that error.
Summary of Contents for 5071A
Page 2: ......
Page 40: ...1 Performance Tests Verifying Specifications 5071A Complete Performance Tests 24 ...
Page 48: ...2 Service Pre Troubleshooting Information 32 Figure 2 1 Instrument Top View ...
Page 168: ...3 Theory of Operation Functional Block Descriptions 152 Figure 3 5 Power Supply Block Diagram ...
Page 173: ......
Page 194: ...5 Replaceable Parts How To Order A Part 178 ...
Page 197: ......