SVAN 106 User MANUAL
_
42
When Pre Trigger position is switched on then the signal is recorded before
the triggering condition with interval, defined by the position Pre of the
Logger Trigger window (path: <Menu> / Measurement / Data Logging).
In the Rec. Limit position it is possible to select the time of event signal
recording after triggering. If the triggering condition appears then the signal
will be recording during the period defined in Rec. Time. The available
values of Rec. Limit are: Max Length, Fixed Len. or Off.
When Off is chosen then the event signal will be recording as long as
memory was filled or until the trigger condition was ceased. When
Max Length is chosen then the signal will be recording during period
defined in the Rec. Time, but can be stopped earlier if trigger condition was
ceased. When Fixed Len. is chosen then the signal will be recording during
period defined in the Rec. Time, even when trigger condition ceased.
In the Rec. Time position it is possible to
select the time of signal recording after
triggering. If next triggering condition
appears then the signal will be recording
additional Rec. Time.
<
4
4
4
4
>
The Channel x positions switch on or off
the channels for event recording.
<
6
6
6
6
>
5.6.5 The marker setup – Marker Setup
The Marker Setup position appears only in
advanced instrument mode (path: <Menu> /
Auxiliary Setup
/
Instrument Mode:
Advanced Mode).
The Marker Setup enables the user to
assign the name for each marker and
define markers for event recording.
<ENT>