SVAN 106 User MANUAL
_
27
5. MEASUREMENT PARAMETERS SETTING – Measurement
The
Measurement
list
contains
the
elements, which enable one to programme
the measurement parameters for all
channels and profiles. The Measurement
list appears after pressing the <Menu>
push-button, selecting the Measurement
text and pressing <ENTER>.
<ENT>
The Measurement list and some of sublist (General Settings and
Data Logging) contents depend on Instrument Mode selection from menu
Auxiliary Setup: Simple Mode or Advanced Mode. In Advanced Mode
some additional functions like triggering, markers, event and wave
recording appear.
Notice: Any parameter in the lists of the Measurement menu
can be changed only when the instrument does not execute a
measurement. The parameters are displayed with different
colour and any marker movement is impossible. The blinking
“play” icon indicates that the instrument is performing the
measurements.
Notice: The parameters can be presented in Logarithm (decibels) or Linear (m/s
2
) units. It
depends on the Scale position value (path: Menu / Display / Results Scale), e.g. 10 m/s
2
can
be presented as 140 dB.
5.1 Selection of measurement parameters - General Settings
The General Settings list consists of the
following parameters: the delay of the start
of
measurements
(Start Delay),
the
integration period (Meas. Period) and the
repetition of the measurement cycles
(Repetition No.). In Advanced menu mode
there are three additional parameters:
Measure Trigger,
Logging Mode
and
Event Recording.