SV106A User Manual
5.11 Programming
the instrument’s internal timer – Timer
The
Timer
position appears only in
Advanced Mode
(
path: <Menu> /
Auxiliary Setup / Instrument Mode
).
Timer
enables the user to programme
the internal real time clock to act as a
delayed start and stop timer. The
instrument
can
be
switched
on
automatically up to 1 month ahead at the
pre-selected
programmed
time
to
perform the measurement with the
same
settings
used
before
the
instrument was switched off.
<ENT>
Selecting the mode of the timer
function
The timer can be switched off (
Off
),
switched on only once (
Single
), or
switched on many times regularly
(
Regular
)
with the period between two
consecutive measurements set in the
Repeat Time
line as 24 hours. It means
that the instrument will be switched on
once a day at the same time until the
user disables the timer function.
<
>
If the instrument is switched on by means of Timer, the “
clock
” icon appears on the screen.
Setting the day for the measurement
to start
The
Start Day
position determines the
date for the measurement to start. The
timer can be programmed up to one
month ahead and during the date setting
the current state of the
R
eal
T
ime
C
lock
is taken into account. The required date
can be selected in the special window,
which is opened by means of the
<
>
,
<
>
push-buttons.
<
>
Setting the time for the
measurement to start
The
Start Time
position determines the
time for the measurement to start. The
required hour and minute can be
selected in a special window, which is
opened by means of the
<
>
,
<
>
push-buttons.
<
>