72
SV106A User Manual
pressing the
<ENTER>
or
<ESC>
push-
button.
Editing the time is performed in the special window, which is opened after
pressing the
<
>
push-button. The selection of the correct parameter
(hour, minute, second, as well as day, month and year) is performed using
the
<
>
,
<
>
push-buttons and the change of its value
– using the
<
>,
<
>
push-buttons pressed together with
<Alt>
.
The required date can be selected in a special window, which is opened
after pressing the
<
>
,
<
>
push-buttons when the
Start Day
text is
displayed inversely in the
Timer
sub-list.
In order to set data, the user has to select its position by means of the
<
>
,
<
>
and
<
>
,
<
>
push button and then press
<ENTER>
to accept the
chosen value.
8.5 Activating the remote-control error confirmation - Remote Control
The
Remote Control
position enables the
user to activate or deactivate error
confirmation
function.
If
the
Remote Control
function is
Enabled
then
the instrument confirms warnings after 5
seconds and the user reaction is not
required. This function is very useful when
the instrument is working as remote
controlled. If the
Remote Control
function
is
Disabled
then the instrument waits for
the user reaction. This mode is used in
normal operation of the instrument.
<ENT>
8.6 Checking the specification of the instrument - Unit Label
The
Unit Label
enables the user to check
the model number of the instrument, its
serial number, the current software version
installed in it and the relevant standards,
which the instrument fulfils.
The displayed text is scrolled on the display
after pressing
<
>
and
<
>
.
<
>
Note: The contents of the Unit Label window should be always sent to Svantek local service
department or official representative in case of any problems faced by the user during the
instrument’s normal operation in the field.