175
SV
104BIS User Manual
D.2
D
EFINITIONS AND FORMULAS OF THE
SLM
FUNCTION RESULTS
The instrument calculates the sound measurement results for three profiles. The calculation flow diagram for
one profile is presented below:
OVL
Percentage of the overloaded input signal, which occurred during the current time period
of the measurement (
T
)
L(A/C/Z)peak
Peak sound level expressed in dB, for
frequency weightings A, C, Z, symbols are
LApeak
,
LCpeak
and
LZpeak
. Peak sound
level is calculated for the given
T
.
( )
=
2
0
2
W
p
t
p
max
log
10
Peak
T
L(A/C/Z)(S/F/I)
max
The highest time weighted sound level (
Max
)
expressed in dB, within a stated time interval,
for frequency weightings A, C, Z and time
weightings F, S, I symbols are
LAFmax
,
LASmax
,
LCFmax
,
LCSmax
etc.
( )
=
2
0
2
W
p
t
p
max
log
10
Max
T
p
2
wτ
(t)
τ
= Fast
τ
= Slow
τ =
Imp
p
w
(t)
p(t)
p
2
w
(t)
L
peak
L
max,
L
min, L,
L
tm3,
L
tm5
L
eq,
LE,
LEPd,
SEL,
L(den),
LEPd,
LR15,
LR60
p
w
(t)
all
DOSE
results
Summary of Contents for SV 104BIS
Page 139: ...139 SV104BIS User Manual 250 Hz 315 Hz 400 Hz 500 Hz...
Page 140: ...140 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Page 141: ...141 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 142: ...142 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 143: ...143 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Page 144: ...144 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 148: ...148 SV104BIS User Manual Figure C 4 Total directional characteristics asymmetrical axis...
Page 150: ...150 SV104BIS User Manual 630 Hz 800 Hz 1000 Hz 1250 Hz...
Page 151: ...151 SV104BIS User Manual 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Page 152: ...152 SV104BIS User Manual 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Page 153: ...153 SV104BIS User Manual 4500 Hz 5000 Hz 5600 Hz 6300 Hz...
Page 154: ...154 SV104BIS User Manual 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Page 166: ...166 SV104BIS User Manual 1 3 octave filters for 16 0 Hz 1 1 octave filter...