To enable an excitation channel click
Excitation On
and then select the channel number to use
and the voltage level to output. Xpert can generate excitation voltages on any channel (1-6).
Filter notch: This is a built in noise filter, the defaults of 60 should work for many
measurements. If AC line noise becomes an issue, lowering this value will help reduce the noise.
Lowering this value will also make measurements take longer.
Settling Time: This is the amount of time in milliseconds to leave the A/D (circuitry, switched 12
volts and VREF to sensors) powered before taking measurements. Sensors requiring a longer
settling time, also called warm-up time, should have this value increased. Range for using the
switched 12 volts is 0 to 30000 (30 seconds) for external AIO module.