Chapter 6: Setting Up Inputs and Outputs
Figure 12: Isolated RS-232 Wiring
Step Two: Configure the Analyzer
The default baud rate for the Analyzer is 19200 and must match that of the system collecting the data. If
you need to change the Analyzer baud rate:
1. Press
SETUP
from the
Main Menu
.
2. Using the
UP
and
DOWN
arrow keys, highlight
INPUT/OUTPUT
and press
ENTER
.
3. Highlight
SERIAL PORT
to access the settings for the Analyzer’s serial port. Using the
UP
and
DOWN
arrow keys, highlight the desired baud rate and press
ENTER
(or press
EXIT
to exit the
SERIAL PORT
menu without saving any changes to the baud rate setting).
4. Press
EXIT
three times to return to the
Main Menu
.
Step Three: Connect the RS-232 Cable to the Computer
Step Four: Cycle the Analyzer Power
The Analyzer must have its power cycled for the serial port baud rate setting to take effect and for the
computer to recognize the Analyzer. To do this:
1. Turn off and then turn on the Analyzer using the power switch on the side of the Analyzer.
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Page 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...