Chapter 5: System Configuration
The Analyzer is shipped with the following defaults:
•
MODE
: Analysis
•
INSTRUMENT LOCATION
: None
•
CLOCK
: Current Date and Time in Boulder, CO USA
• I
NPUT/OUPUT
•
BINARY INPUTS
: Disabled
•
AC FREQUENCY
: 60Hz
•
ANALOGS OUTPUTS
: No Output
•
ALARM OUTPUTS
: Off
•
SERIAL OUTPUT
: 19200 Baud Rate
•
BINARY OUTPUTS
: Off
•
PRINTER TYPE
: Seiko DPU-414
• STREAM SETUP
•
FLUSH TIME
: 2 minutes
•
SAMPLE METHOD
: Individual
•
ANALYSIS TIME
: 12 minutes
• STREAM SETTINGS
•
STREAM 1
: Enabled; 10 Samples/Stream
•
STREAM 2, 3, 4
: Disabled
MODE
There are five modes (operations) available:
ANALYSIS, CELL SLOPE CAL, CELL OFFSET CAL,
CALIBRATION
, and
VERIFICATION
.
• The
ANALYSIS
mode measures boron in on-line samples and is the default mode. To start
analysis, press
START
from the
Main Menu
. Once started, this mode will run continuously until
commanded to stop (by pressing
STOP
).
If you need to return the Analyzer to the
ANALYSIS
mode:
1. Press
SETUP
from the
Main Menu
.
2. Highlight
MODE
and press
ENTER
.
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...
Page 104: ...Notes SUEZ 2018 104 of 104 DLM 71000 03 EN Rev B 104...