![Suez Sievers UPW Boron Analyzer Operation And Maintenance Manual Download Page 53](http://html1.mh-extra.com/html/suez/sievers-upw-boron-analyzer/sievers-upw-boron-analyzer_operation-and-maintenance-manual_1389722053.webp)
SUEZ © 2018
DLM 71000-03 EN
Rev. B
Chapter 5: System Configuration
6. Press the
key to print the values displayed on the screen.
7. Press
EXIT
once to return to the previous menu or three times to return to the
Main Menu
.
To display the data in a graphical format:
1. Press
ENTER
while the
BORON GRAPH
option is highlighted to display the data in a graphical
format. The most-recent data is displayed at the right.
• The
DATE
and
TIME
stamps of the data are not displayed on the graph; instead, the
DATE
and
TIME
of the right-most data point is displayed.
• The
OFFSET
, in hours or days, from that point is displayed on the x-axis.
2. The
SCALE
key allows the x- and y-axis scales to be changed. Pressing the
SCALE
key makes
available the following keys:
TIME
, which changes the x-axis scale;
YBASE
, which changes the
lower bound of the y-axis; and
YTOP
, which changes the upper bound of the y-axis.
• Selecting
TIME
will allow the x-axis to be set back to the default (7-day) scale, or allow it to be
set to any of the preselected values by using the
UP
and
DOWN
arrow keys. Pressing
ENTER
will cause the new scale to take effect; pressing
EXIT
will revert to the previous scale value.
• The
YTOP
and
YBASE
y-axis scale adjustments work in an identical manner, with the
exception that the default value is selected by pressing the
AUTO
key. The
AUTO
key
calculates the scale adjustment, based upon currently displayed data, to best fit on the screen.
The
LEFT
arrow key advances data, based upon the x-axis scaling, to an earlier point in the
total time graphed. The
RIGHT
arrow key advances data, based upon the x-axis scaling, to a
later point in the total time graphed. A time difference between two data points of more than
15 minutes is indicated on the display as a vertical line. If the data is viewed during analysis, a
newly acquired data point will update the screen at the right-most position and the countdown
time will appear in the upper right corner.
3. Press
to print the graph.
4. Press
EXIT
once to return to the previous menu or three times to return to the
Main Menu
.
CAL/VERIFY HIST
The
CAL/VERIFY HIST
option allows viewing of the calibration/verification history of the Analyzer.
1. Press
VIEW
from the
Main Menu
to see the calibration/verification history of the Analyzer.
2. Highlight
CAL/VERIFY HIST
by using the arrow keys and press
ENTER
.
• The history data is displayed with the most recent data being displayed first.
• The date of the calibration/verification is displayed with a code indicating what type of
calibration/verification was performed.
• To view the specific data of one of the calibration/verifications, use the
UP
and
DOWN
arrow
keys to highlight the desired record, then press VIEW to see the data for that calibration or
verification.
Summary of Contents for Sievers UPW Boron Analyzer
Page 4: ...SUEZ 2018 4 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 16: ...SUEZ 2018 16 of 104 DLM 71000 03 EN Rev B THIS PAGE IS INTENTIONALLY LEFT BLANK...
Page 103: ...Notes SUEZ 2018 103 of 104 DLM 71000 03 EN Rev B 104...