56
S
e. Run wires thru the HCT down the horizontal arm and out the access of the contact block.
f. Replace the 5mm hex screw on the side of the HCT and tighten.
g. Proceed to step 8.
7. For units with no HCT
a. Remove upper front blue cover by removing the 2 3mm screws.
b. Pull the cover up and carefully remove the wireset from the arm.
c. Disconnect the wireset from bracket by removing 2 Torx Ejot screws.
d. Attach new wireset to bracket using the 2 screws removed.
e. Run wires thru the horizontal arm and out the access of the contact block.
f. Replace the upper front blue cover by reinstalling the 2 3mm screws and tighten.
8. Attach wireset to contact block and tape the 4 pin connector to ensure it does not disconnect.
9. Place the contact block into the suspension and verify that all prongs are touching the slip ring(see
above)
10. Place the plastic cover into place, making sure that the connection end of the cover snaps into place on
the contact block.
CAUTION: If the cover is not snapped into place correctly, the suspension will not
function properly.
11. Replace the 2 cross head screws and tighten.
12. Reinstall Spring Arm per section 7.5.
13. Reinstall Light Head per section 7.9.
14. Restore power and check the system for proper operation.