121
S
• Remove existing “N” label from the existing
neutral block for reuse.
• Using service kit P41494:
• Connect the 8 amp fuse (500393) to the
fuse clamp (39965) and secure by twisting
clockwise. Add end bar (39967) to the fuse
clamp.
• Mount the new Line, Neutral, and Ground
blocks to the DIN rail as shown.
• Add label “N” to the terminal block. Add label “L”
to the fuse clamp.
• Repeat for all lights in the suspension.
• Reconnect Chromophare system wires to the
new blocks: black wire to Line block (fuse clamp),
white wire to Neutral block (not fused), each
green and yellow wire to a ground block.
• An electrician is required to reconnect hospital
mains to the terminal blocks and restore power to
the room.
SK Plate
For Tube Mounted electronics:
• Remove the current clamps from the mounting rail for the affected light by pinching the front clip for
easy release.
• For tube mounted electronics, the line, neutral, and ground blocks must all be replaced for all
lights on the suspension.
• Remove existing “N” label from the existing neutral block for reuse.