5. To install the dome:
(a) Fit new gasket onto edge of frame (if needed), making sure the gasket is not stretched. To keep the
gasket from stretching, push back on the gasket every 3-4 inches until the complete unit is installed.
(b) If present, connect the EndoLite wiring back to CPU connector J8.
(c) Place the dome on top of the light and fit into gasket on one side of the light.
(d) Place pressure on the side of the dome pulling toward the portion that is in the gasket.
(e) Slide the gasket tool under the side of the gasket and work around until gasket is sealed on the
entire dome.
9.4.2 EndoLite
The EndoLite is an optional 20 watt halogen light mounted to the center of the dome on E HID and Halogen
series lights.
Summary of Contents for Berchtold CHROMOPHARE