114
S
b)
Slide the DIP switch no. 3 to ON position as shown below.
Plug in the power / CAN connector and wait until the program “TSLIB calibration utility” starts on the screen.
c) Calibrate the touch controller by pressing the finger on the crosshair positions until the program
“TSLIB test program” starts on the screen.
d) Unplug the unit and slide the DIP switch no. 3 to OFF position and reassemble the unit.
9. Touch Screen Addressing – Cardanic Controls
a) Turn the light on using the rocker switch on the light head frame.
b) To activate the service function, hold down the hidden service field for approximately 3 seconds.
Hidden service field lower left hand corner
Summary of Contents for Berchtold CHROMOPHARE
Page 1: ...CHROMOPHARE Surgical Lights Single and Combination Lights Service and Parts Manual ...
Page 2: ......
Page 64: ... 59 S ...
Page 65: ...60 S ...