55
S
7.17.2
Replacing internal wiring on Horizontal Arm (light suspension only)
1. Secure power to the system from the circuit breaker.
2. Remove Light Head from Spring Arm per section 7.9.
3. Remove Spring Arm per section 7.5.
4. Remove 2 cross head screws on the bottom of the horizontal arm next to the central shaft to access the
contact block.
5. Carefully remove contact block with the wire set attached. Disconnect wireset from contact block.
Note: Contact block is fragile and can be damaged if care is not observed.
6. For units with HCT
a. Remove the 5mm hex screw on the side of the HCT,
b. Remove wireset and bracket from the arm.
c. Disconnect wireset from bracket by removing the 2 4mm screws.
d. Attach new wireset to bracket using the 2 screws removed.
Chromophare suspension wiring with HCT